4.6 Article

Assay for the simultaneous determination of acetaminophen-caffeine-butalbital in human serum using a monolithic column

Journal

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
Volume 36, Issue 4, Pages 737-741

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpba.2004.07.042

Keywords

monolithic column; acetaminophen-caffeine-butalbital; serum; HPLC

Ask authors/readers for more resources

A fast and sensitive high performance liquid chromatography (HPLC) assay was developed on a C18 monolithic column for the simultaneous determination of acetaminophen-caffeine-butalbital in human serum. Serum samples were treated with a solid phase extraction procedure. The analytes were separated using a mobile phase of 95:5 (v/v) 0.1 M potassium phosphate monobasic (pH 2.41)-acetonitrile on the C 18 monolithic column with detection at 220 nm. Benzoic acid was used as the internal standard (IS). The method was validated over the range of 1.25-100 mug/ml for each drug and found to be linear (r > 0.995, n = 12) with RSD less than 8.3%. The method proved to be accurate (percent bias for all calibration samples varied from - 14.6 to - 1.3%) and precise (ranged from 2.9 to 13.4%). The mean percent absolute recoveries from serum were 89.7 +/-3.6 for acetaminophen, 95.5 +/- 4.5 for caffeine, 99 +/- 5.2 for butalbital and 83.4 +/- 3.9% for the internal standard. (C) 2004 Elsevier B.V. All rights reserved.