4.6 Article

Poly(vinyl alcohol) cryogel phantoms for use in ultrasound and MR imaging

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 49, Issue 24, Pages 5529-5546

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/49/24/009

Keywords

-

Ask authors/readers for more resources

Poly(vinyl alcohol) cryogel, PVA-C, is presented as a tissue-mimicking material, suitable for application in magnetic resonance (MR) imaging and ultrasound imaging. A 10% by weight poly(vinyl alcohol) in water solution was used to form PVA-C, which is solidified through a freeze-thaw process. The number of freeze-thaw cycles affects the properties of the material. The ultrasound MR Imaging characteristic were investigated using cylindrical samples of PVA-C. The speed of sound was found to range 1520 to 1540 m s(-1), and the attenuation coefficients were in the range of 0.075-0.28 dB (cm MHz)(-1). T1 and T2 relaxation values were found to be 718-1034 ms and 108-175 ms, respectively. We also present applications of this material in an anthropomorphic brain phantom, a multi-volume stenosed vessel phantom and breast biopsy phantoms. Some suggestions are made for how best to handle this material in the phantom design and development process.