Journal
ANALYTICA CHIMICA ACTA
Volume 529, Issue 1-2, Pages 123-127Publisher
ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2004.10.054
Keywords
generic sulfonamides assay; screening test; SPR biosensor; pork muscle; sulfamethazine sulfadiazine
Categories
Ask authors/readers for more resources
A binding protein displaying broad-spectrum cross-reactivity within the sulfonamide group was used in conjunction with a sulfonamide specific sensor chip and a surface plasmon resonance biosensor to develop a rapid broad spectrum screening assay for sulfonamides in porcine muscle. Results for 40 samples were available in just over 5 h after the completion of a simple sample preparation protocol. Twenty sulfonamide compounds were detected. Acetylated metabolites were not recognised by the binding protein. Limit of detection (mean-three times standard deviation value when n = 20) was calculated to be 16.9 ng g(-1) in tissue samples. Intra-assay precision (n = 10) was calculated at 4.3 %CV for a sample spiked at 50 ng g(-1) with sulfamethazine, 3.6 %CV for a sample spiked at 100 ng g(-1) with sulfamethazine, 7.2 %CV for a sample spiked at 50 ng g(-1) with sulfadiazine and 3.1 %CV for a sample spiked at 100 ng g-1 with sulfadiazine. Inter-assay precision (n = 3) was calculated at 9.7 %CV for a sample spiked at 50 ng g-1 with sulfamethazine, 3.8 %CV for a sample spiked at 100 ng g(-1) with sulfamethazine, 3.5 %CV for a sample spiked at 50 ng g(-1) with sulfadiazine and 2.8 %CV for a sample spiked at 100 ng g(-1) with sulfadiazine. (C) 2004 Elsevier B.V. All rights reserved.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available