Measurement of macular pigment: Raman spectroscopy versus heterochromatic flicker photometry

Journal

INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE
Volume 46, Issue 3, Pages 1023-1032

Publisher

ASSOC RESEARCH VISION OPHTHALMOLOGY INC
DOI: 10.1167/iovs.04-1032

Keywords

-

Categories

Ask authors/readers for more resources

PURPOSE. There are several techniques for measuring macular pigment (MP) in vivo, of which Raman spectroscopy (RS) is a recently developed objective method. This study reports the reproducibility, test-retest variability, and validity of RS MP readings, by comparing them with heterochromatic flicker photometry (HFP). METHODS. MP was measured with HFP and RS in 120 healthy subjects, and the latter technique was also used on two separate occasions in a sample of 20 subjects to investigate the intersessional variability of readings. Intrasessional reproducibility of RS MP measurements was also calculated. In addition, serum concentrations of lutein (L) and zeaxanthin (Z) were measured and correlated with both RS and HFP NIP readings. RESULTS. Mean (+/-SD) NIP in the right eye was 0.279 +/- 0.145 and 0.319 +/- 0.155 with RS and HFP, respectively. The differences between corresponding MP readings taken on RS and HFP lay within the Bland-Altman 95% limits of agreement for the two instruments in 93.6% and 94.4% of cases in the right and left eyes, respectively. Intrasessional reproducibility of RS readings, expressed as the coefficient of variation, was 8.42% +/- 7.12%. Ninety-five percent of MP readings taken with RS on two separate occasions lay within the 95% limits of agreement for the two sessions. A positive, but insignificant, relationship was observed between RS and HFP MP readings and serum concentrations of L and Z (RS, P = 0.356; HFP, P = 0.540). CONCLUSIONS. RS, an objective method of measuring NIP levels in vivo, exhibits acceptable reproducibility and test-retest variability. The results demonstrated good correlation between RS and HFP measurements of MP, thus authenticating RS against a validated psychophysical technique of measuring MP. However, investigators should use only one of these instruments for the duration of any given study because of differences in the scientific rationale, and the factors that influence RS and HFP measurements of MP.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available