4.6 Article Proceedings Paper

On calibration of pH meters

Journal

SENSORS
Volume 5, Issue 4-5, Pages 209-219

Publisher

MDPI
DOI: 10.3390/s5040209

Keywords

electrochemistry; electrode; pH measurement

Ask authors/readers for more resources

The calibration of pH meters including the pH glass electrode, ISE electrodes, buffers, and the general background for calibration are reviewed. Understanding of basic concepts of pH, pOH, and electrode mechanism is emphasized. New concepts of pH, pOH, as well as critical examination of activity, and activity coefficients are given. The emergence of new solid state pH electrodes and replacement of the salt bridge with a conducting wire have opened up a new horizon for pH measurements. A pH buffer solution with a conducting wire may be used as a stable reference electrode. The misleading unlimited linear Nernstian slope should be discarded. Calibration curves with 3 nonlinear portions for the entire 0 - 14 pH range due to the isoelectric point change effect are explained. The potential measurement with stirring or unstirring and effects by double layer (DL) and triple layer (TL) will be discussed.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available