4.7 Article

Determination of camptothecin and 10-hydroxycamptothecin in human plasma using polymer monolithic in-tube solid phase microextraction combined with high-performance liquid chromatography

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 382, Issue 1, Pages 204-210

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-005-3194-4

Keywords

camptothecin; 10-hydroxycamptothecin; high performance liquid chromatography; in-tube solid-phase microextraction; poly(methacrylic acid-ethylene glycol dimethacrylate) monolithic capillary

Ask authors/readers for more resources

A biocompatible in-tube solid-phase microextraction (SPME) device was used for the direct and on-line extraction of camptothecin and 10-hydroxycamptothecin in human plasma. Biocompatibility was achieved through the use of a poly(methacrylic acidethylene glycol dimethacrylate) monolithic capillary column for extraction. Coupled to high performance liquid chromatography (HPLC) with UV detection, this on-line in-tube SPME method was successfully applied to the simultaneous determination of camptothecin and 10-hydroxycamptothecin in human plasma. The calculated detection limits for camptothecin and 10-hydroxycamptothecin were found to be 2.62 and 1.79 ng/mL, respectively. The method was linear over the range of 10-1000 ng/mL. Excellent method reproducibility was achieved, yielding RSDs of 2.49 and 1.59%, respectively. The detection limit (S/N = 3) of camptothecin was found to reach 0.1 ng/mL using fluorescence detection. The proposed method was shown to cope robustly with the extraction and analysis of camptothecin and 10-hydroxycamptothecin in plasma samples.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available