4.6 Article Proceedings Paper

Dual-frequency ultrasound for detecting and sizing bubbles

Journal

ACTA ASTRONAUTICA
Volume 56, Issue 9-12, Pages 1041-1047

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.actaastro.2005.01.032

Keywords

-

Ask authors/readers for more resources

ISS construction and Mars exploration require extensive extravehicular activity (EVA), exposing crewmembers to increased decompression sickness risk. Improved bubble detection technologies could help increase EVA efficiency and safety. Create Inc. has developed a bubble detection and sizing instrument using dual-frequency ultrasound. The device emits pump and image signals at two frequencies. The low-frequency pump signal causes an appropriately-sized bubble to resonate. When the image frequency hits a resonating bubble, mixing signals are returned at the sum and difference of the two frequencies. To test the feasibility of transcutaneous intravascular detection, intravascular bubbles in anesthetized swine were produced using agitated saline and decompression stress. Ultrasonic transducers on the chest provided the two frequencies. Mixing signals were detected transthoracically in the right atrium using both methods. A histogram of estimated bubble sizes could be constructed. Bubbles can be detected and sized transthoracically in the right atrium using dual-frequency ultrasound. © 2005 Elsevier Ltd. All rights reserved.