4.5 Article

Laminar flow resistance in short microtubes

Journal

INTERNATIONAL JOURNAL OF HEAT AND FLUID FLOW
Volume 26, Issue 3, Pages 506-512

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ijheatfluidflow.2004.11.002

Keywords

-

Ask authors/readers for more resources

We have measured the pressure drop for the flow of liquid through a series of short microtubes ranging from 80 to 150 pm in diameter with aspect ratios between L/D = 2 and L/D = 5. These dimensions were selected to resemble lumens of implantable microstents that are under consideration for the treatment of glaucoma. For physiologically relevant pressure drops and flow rates, we have determined that a fully-developed laminar pipe flow may be assumed throughout the microtube when (L/D) > 0.20Re, where Re is the Reynolds number based on the diameter, D, and L is the length of the tube. We have examined flow rates between 0.1 and 10 mu L/s, corresponding to Reynolds numbers between 1 and 150. For smooth microtubes, no difference from macroscopic flow is observed for the tube sizes considered. However, flow resistance is found to be sensitive to the relative surface roughness of the tube walls. (c) 2004 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available