4.6 Article

Precise radiochromic film dosimetry using a flat-bed document scanner

Journal

MEDICAL PHYSICS
Volume 32, Issue 7, Pages 2245-2253

Publisher

WILEY
DOI: 10.1118/1.1929253

Keywords

radiochromic film dosimetry; flat-bed scanner; uncertainty analysis; sensitivity curves

Ask authors/readers for more resources

In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic(R) film models, HS and Prototype A EBT exposed to 6 MV photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around 0.4 Gy in the case of the EBT (Prototype A), and for doses above 5 Gy in the case of the HS model GafChromic(R) film using a region of interest 2 X 2 mm(2) in size. (C) 2005 American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available