4.5 Article

Computer-based detection and analysis of heart sound and murmur

Journal

ANNALS OF BIOMEDICAL ENGINEERING
Volume 33, Issue 7, Pages 937-942

Publisher

SPRINGER
DOI: 10.1007/s10439-005-4053-3

Keywords

detection algorithm; signal analysis; short-time Fourier transform; systolic murmur

Ask authors/readers for more resources

To develop a digital algorithm that detects first and second heart sounds, defines the systole and diastole, and characterises the systolic murmur. Heart sounds were recorded in 300 children with a cardiac murmur, using an electronic stethoscope. A Digital algorithm was developed for detection of first and second heart sounds. R-waves and T-waves in the electrocardiography were used as references for detection. The sound signal analysis was carried out using the short-time Fourier transform. The first heart sound detection rate, with reference to the R-wave, was 100% within 0.05-0.2R-R interval. The second heart sound detection rate between the end of the T-wave and the 0.6R-R interval was 97%. The systolic and diastolic phases of the cardiac cycle could be identified. Because of the overlap between heart sounds and murmur a systolic segment between the first and second heart sounds (20-70%) was selected for murmur analysis. The maximum intensity of the systolic murmur, its average frequency, and the mean spectral power were quantified. The frequency at the point with the highest sound intensity in the spectrum and its time from the first heart sound, the highest frequency, and frequency range were also determined. This method will serve as the foundation for computer-based detection of heart sounds and the characterisation of cardiac murmurs.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available