4.6 Article

Estimation of functional residual capacity at the bedside using standard monitoring equipment:: A modified nitrogen washout/washin technique requiring a small change of the inspired oxygen fraction

Journal

ANESTHESIA AND ANALGESIA
Volume 101, Issue 1, Pages 206-212

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1213/01.ANE.0000165823.90368.55

Keywords

-

Categories

Ask authors/readers for more resources

We developed a modified nitrogen washin/washout technique based on standard monitors using inspiratory and end-tidal gas concentration values for functional residual capacity (FRC) measurements in patients with acute respiratory failure (ARF). For validation we used an oxygen-consuming lung model ventilated with an inspiratory oxygen fraction (FIO2) between 0.3 and 1.0. The respiratory quotient of the lung model was varied between 0.7 and 1.0. Measurements were performed changing FIO2 with fractions of 0.1, 0.2, and 0.3. In 28 patients with ARF, duplicate measurements were performed. In the lung model, an FIO2 change of 0.1 resulted in a value of 103 +/- 5% of the reference FRC value of the lung model, and the precision was equally good up to an FIO2 of 1.0 with a value of 103 +/- 7%. In the patients, duplicate measurements showed a bias of - 5 mL with a 95% confidence interval [-38; 29 mL]. A comparison of a change in FIO2 of 0.1 with 0.3 showed a bias of -9 mL and limits of agreement of [ - 365; 347 mL]. This study shows good precision of FRC measurements with standard monitors using a change in 17102 of only 0.1. Measurements can be performed with equal precision up to an FIO2 of 1.0.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available