Purpose: To assess 3-T imaging of the knee. Materials and methods: We reviewed 357 3-T magnetic resonance images of the knee obtained using a dedicated knee coil. From 58 patients who had arthroscopy we determined the sensitivity and specificity for anterior cruciate ligament (ACL) tear and medial and lateral meniscal tear. Results: A chemical shift artifact showed prominently at 3 T even after improvements had been made by increasing the bandwidth. For complete ACL tear the sensitivity was 100% (95% confidence interval, CI, 75.30-100.00), and the specificity was 97.9% (95% CI 87.7-99.9). For the medial meniscus the sensitivity was 100.00% (95% CI 90.0-100.00), and the specificity was 83.3%(95% CI 66.6-95.3). For the lateral meniscus the sensitivity was 66.7% (95% CI 38.4-88.2), and the specificity was 97.6% (95% CI 87.1-99.9). Conclusions: In general 3-T imaging allows a favorable display of anatomy and pathology. The lateral meniscus was assessed to be weaker than the other anatomic structures. Three-tesla imaging allows increased signal-to-noise ratio, increased resolution, and faster scanning times.