4.2 Article

Evaluation of the extraction method for the cytotoxicity testing of latex gloves

Journal

YONSEI MEDICAL JOURNAL
Volume 46, Issue 4, Pages 579-583

Publisher

YONSEI UNIV COLL MEDICINE
DOI: 10.3349/ymj.2005.46.4.579

Keywords

extraction conditions; cytotoxicity; latex gloves; L-929 cells; medical devices

Ask authors/readers for more resources

In this study, the cytotoxicity of medical latex gloves to cultured L-929 cells was determined using various extraction conditions. According to the extraction time and temperature, three types of extraction conditions were used: 1) 24 h at 37 degrees C; 2) 72 h at 37 degrees C; 3) 72 h at 50 degrees C. Also, four different extraction vehicles were used, namely, distilled water (DW), 9 g/l sodium chloride (saline) in DW, and culture media with or without serum. Under the above-mentioned conditions, the samples were extracted and then 2-fold serially diluted in the concentration range 3.13-50%. When extracted with either DW or saline for 24 h or 72 h at 37 degrees C, only 50% diluted samples showed distinct cytotoxicity to L-929 cells. Moreover, no cytotoxic potentials were observed when gloves were extracted with DW or saline at 50 degrees C for 72 h. Cytotoxicity was markedly greater when gloves were extracted with culture medium, irrespective of the presence of serum in the medium. These results suggest that optimal extraction conditions should be established for the cytotoxicity evaluations of biomaterials and medical devices.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available