A G4.0 poly(propyleneimine) dendrimer immobilized silica gel (PPID-SG) was prepared with a divergent approach from gamma-aminopropyl silica gel (APSG) by repeating the reactions: 1) cyanoethylation of -NH2 and 2) reduction of -CN. Then it was successfully applied as the microcolumn packing for the on-line flow-injection (FI) preconcentration and separation and flame atomic absorption spectrometry (FAAS) determination of Pt. A limit of detection (LOD) of 0.014 mu g/ml, was obtained when 0.5 mu g/mL Pt in chloride acid of PH = 3 was preconcentrated with a sample flow rate of 4.6 mL/min for 60s and the relative standard deviation (RSD) was 2.65%. Coexisting metal ions with concentration of 5.0 mg/mL and anion ions of 20.0 mg/ml have no interference in the determination of Pt. The proposed method was successfully applied to the determination of Pt in nickel alloy and anode slime samples.