4.5 Article

Image reconstruction in SNR units: A general method for SNR measurement

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 54, Issue 6, Pages 1439-1447

Publisher

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.20713

Keywords

MRI; SNR measurement; SENSE; parallel MRI; phased array; noise

Funding

  1. Intramural NIH HHS [Z01 HL004608-08] Funding Source: Medline

Ask authors/readers for more resources

The method for phased array image reconstruction of uniform noise images may be used in conjunction with proper image scaling as a means of reconstructing images directly in SNR units. This facilitates accurate and precise SNR measurement on a per pixel basis. This method is applicable to root-sum-of-squares magnitude combining, B-1-weighted combining, and parallel imaging such as SENSE. A procedure for image reconstruction and scaling is presented, and the method for SNR measurement is validated with phantom data. Alternative methods that rely on noise only regions are not appropriate for parallel imaging where the noise level is highly variable across the field-of -view. The purpose of this article is to provide a nuts and bolts procedure for calculating scale factors used for reconstructing images directly in SNR units. The procedure includes scaling for noise equivalent bandwidth of digital receivers, FFTs and associated window functions (raw data filters), and array combining. Magn Reson Med 54:1439-1447, 2005. Published 2005 Wiley-Liss, Inc.dagger

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available