Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement

Journal

BRITISH JOURNAL OF OPHTHALMOLOGY
Volume 89, Issue 12, Pages 1581-1585

Publisher

BMJ PUBLISHING GROUP
DOI: 10.1136/bjo.2005.069815

Keywords

-

Categories

Ask authors/readers for more resources

Aims: To compare the values measured for retinal macular thickness with the first and last generations of the optical coherence tomograph (OCT1 and Stratus OCT, Zeiss, Humphrey Division). Methods: This was a cohort study. 59 eyes were examined: 17 had a normal macula and 42 had a diabetic macular oedema. In each eye, mean retinal thickness (RT) was measured automatically in the nine macular Early Treatment Diabetic Retinopathy Study areas and at the foveal centre, using OCT1 and Stratus OCT. The paired mean RT values for each area and the type and proportion of artefacts were compared. Results: Of the 590 automatic measurements, 505 had no artefact, either with OCT1 or Stratus OCT. The mean difference between the OCT1 and Stratus OCT measurements was 25 (SD 26.2) mu m (p < 0.0001). With Stratus OCT, RT values were significantly higher, by 8.1% (7.8%), than with OCT1. Artefacts were only observed in cases of diabetic macular oedema and were significantly more frequent with OCT1 than Stratus OCT (10.5% versus 4.4, p < 0.0001). Conclusion: The macular retinal thickness values measured with Stratus OCT were significantly higher than those measured with OCT1. Stratus OCT has the advantage of producing fewer artefacts than OCT1 in pathological cases.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available