4.6 Editorial Material

A New Simple and Objective Method for Graft Sizing in Valve-Sparing Root Replacement Using the Reimplantation Technique

Journal

ANNALS OF THORACIC SURGERY
Volume 92, Issue 2, Pages 749-751

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.athoracsur.2011.03.015

Keywords

-

Ask authors/readers for more resources

The methods of graft sizing in valve-sparing surgery are criticized for their complexity and subjectivity in application. We propose a simple method for graft sizing in valve-sparing root replacement using the reimplantation technique. Practically, the height of the commissure between the noncoronary cusp and the left coronary cusp give the size of the graft. This new method of graft sizing was successfully applied in the last 27 consecutive patients with good immediate results. Graft sizing with this objective and reproducible simple method results in restoration of normal aortic valve geometry and function. (Ann Thorac Surg 2011;92:749-51) (C) 2011 by The Society of Thoracic Surgeons