Open-path ozone detection by quantum-cascade laser

Journal

APPLIED PHYSICS B-LASERS AND OPTICS
Volume 82, Issue 3, Pages 501-506

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00340-005-2010-3

Keywords

-

Ask authors/readers for more resources

Open-path ozone measurements performed by mid-IR differential absorption spectroscopy are reported. Ozone spectrum was taken by fast repetitive sweeping of a quantum-cascade laser wavelength over a spectral feature from the nu(3) absorption band of ozone, centered at 1031.2 cm(-1). Short (100 ns) sweeping times were essential to prevent line-distortions caused by atmospheric turbulence. For fast wavelength sweeping, a technique that employed the thermal chirp during 140 ns excitation pulses was used. The lowest detection limit of 0.3 ppm.m was estimated from the minimum detectable differential absorption. We present the results from cell and open-path measurements over 440 and 5800 m, together with experimental data regarding the tuning range, the tuning rate and the tuning linearity of the QCL while operated with 140 ns excitation pulses.