Early opacification of silicone intraocular lenses: Laboratory analyses of 6 explants

Journal

JOURNAL OF CATARACT AND REFRACTIVE SURGERY
Volume 32, Issue 3, Pages 499-509

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcrs.2005.12.137

Keywords

-

Ask authors/readers for more resources

PURPOSE: To report laboratory analyses of 6 intraocular lenses (IOLs) explanted from patients who had visual disturbances caused by early postoperative opacification of the lens optic. SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA. METHODS: Six patients with 3-piece silicone lenses presented with optic cloudiness as early as a few hours after implantation. The lenses were implanted in 4 locations in Brazil and in France. The lenses in Brazil were stored at the same location before implantation. Gross and microscopic analyses were performed (dry and hydrated states). One half of each specimen underwent gas chromatography/mass spectrometry (GC/MS) analysis and/or extraction by isopropyl alcohol or acetonitrile. One lens also underwent scanning electron microscopy (SEM) with energy dispersive x-ray spectroscopy (EDS). The IOLs were examined for the presence of contaminants or deposits that could cause fast optic opacification. RESULTS: The IOLs showed a white optic discoloration in the hydrated state but became transparent on complete dehydration. Suspect exogenous chemical compounds were identified in GC/MS analyses; general classes included terpenes and ketones, typically found in industrial cleaning agents and fumigants. Surface analysis (SEM and EDS) did not show any significant deposits on the external surfaces and sagittal cut in I of the specimens. CONCLUSIONS: Most IOLs are enclosed in semipermeable packages to allow sterilization by ethylene oxide gas. During cleaning or disinfection of storage rooms, aerosolized solutions may introduce chemicals through the package and onto the IOLs. This may cause surface changes in the IOL, promoting opacification by water ingress in the aqueous environment. Cleaning and disinfection procedures of IOL storage areas should be monitored carefully.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available