Comparison of temperature changes in the pulp chamber induced by various light curing units, in vitro

Journal

OPERATIVE DENTISTRY
Volume 31, Issue 2, Pages 261-265

Publisher

OPERATIVE DENTISTRY INC
DOI: 10.2341/05-26

Keywords

-

Ask authors/readers for more resources

This study compared the temperature increase in a pulp chamber as a result of using various light-curing units during resin composite polymerization, and it determined the effect of remaining dentin thickness on temperature rise. A Class II occlusodistal cavity with a remaining dentin thickness of 2 nun was prepared in an extracted human mandibular molar. A 2-mm layer of fine hybrid resin composite was placed on the floor of the proximal box. A K-type thermocouple was inserted into pulp chambers filled with heat sink compound, and pulp chamber temperature rise (starting temperature: 37.0 +/- 0.1 degrees C) during polymerization of the composite was measured. The light-curing units tested included two halogen lights, Spectrum 800 and Elipar Trilight (Standard and Exponential mode); a light-emitting diode (LED, Elipar Freelight) and a plasma arc (Virtuoso, Xenon Power Arc). Irradiation time was 40 seconds for the halogen and LED lights and 3 seconds for the plasma arc light. Five measurements were carried out for every light-curing unit. The same experimental design was conducted after the cavity preparation was modified, leaving a 1-mm. thick dentin layer. The Kruskal-Wallis and multiple comparison tests were used to evaluate the differences among the tested curing units. Mann Whitney-U tests were used to compare the mean temperature rise in each curing unit for different remaining dentin thicknesses. The increase in pulp chamber temperature ranged between 1.40-3.8 degrees C. The highest temperature rise was observed when using Elipar Trilight Standard mode, and the lowest temperature rise was observed with light emitting diode for both remaining dentin thicknesses. The only significant differences in temperature rise were observed between Elipar Trilight Standard mode and LED. No significant difference (p > 0.01) existed for the different modes of Elipar Trilight. A statistically significant higher temperature rise was observed within each curing unit at a depth of 1 m, compared to 2 mm. Although the tested light-curing units caused a temperature rise in the pulp chamber, none exceed the critical value of 5.5 degrees C.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available