3.9 Article

A comparison of assay performance measures in screening assays:: Signal window, Z′ factor, and assay variability ratio

Journal

JOURNAL OF BIOMOLECULAR SCREENING
Volume 11, Issue 3, Pages 247-252

Publisher

SAGE PUBLICATIONS INC
DOI: 10.1177/1087057105285610

Keywords

signal window; Z ' factor; assay variability ratio; assay validation

Ask authors/readers for more resources

In this article, the authors compare the assay performance measures, signal window, Z' factor, and assay variability ratio. They exanl ine their mathematical formulae for similarities and differences, describe their statistical sampling properties using the results of a Computer simulation, and illustrate their use with example data. Based on these results, the authors recommend the Z' factor as a preferred measure of assay performance for screening assays and point out that none of these measures are adequate for characterizing concentration-response assays.