Scatter rejection in multislit digital mammography

Journal

MEDICAL PHYSICS
Volume 33, Issue 4, Pages 933-940

Publisher

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2179122

Keywords

mammography; scatter; scanning; slit; slot

Ask authors/readers for more resources

The scatter to primary ratio (SPR) was measured on a scanning multislit full-field digital mammography system for different thickness of breast equivalent material and different tube voltages. Scatter within the detector was measured separately and was found to be the major source of scatter in the assembly. Measured total SPRs below 6% are reported for breast range 3-7 cm. The performance of the multislit assembly is compared to other imaging geometries with different scatter rejection schemes by using the scatter detective quantum efficiency. (c) 2006 American Association of Physicists in Medicine.