Intensity non-uniformity correction in MRI: Existing methods and their validation

Journal

MEDICAL IMAGE ANALYSIS
Volume 10, Issue 2, Pages 234-246

Publisher

ELSEVIER
DOI: 10.1016/j.media.2005.09.004

Keywords

magnetic resonance imaging; intensity non-uniformity artifact; image intensity correction; validation methods

Ask authors/readers for more resources

Magnetic resonance imaging is a popular and powerful non-invasive imaging technique. Automated analysis has become mandatory to efficiently cope with the large amount of data generated using this modality. However, several artifacts, such as intensity nonuniformity, can degrade the quality of acquired data. Intensity non-uniformity consists in anatomically irrelevant intensity variation throughout data. It can be induced by the choice of the radio-frequency coil, the acquisition pulse sequence and by the nature and geometry of the sample itself. Numerous methods have been proposed to correct this artifact. In this paper, we propose an overview of existing methods. We first sort them according to their location in the acquisition/processing pipeline. Sorting is then refined based on the assumptions those methods rely on. Next, we present the validation protocols used to evaluate these different correction schemes both from a qualitative and a quantitative point of view. Finally, availability and usability of the presented methods is discussed. (C) 2005 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available