4.5 Article

Criterion validity study of the cervical range of motion (CROM) device for rotational range of motion on healthy adults

Journal

JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY
Volume 36, Issue 4, Pages 242-248

Publisher

J O S P T,
DOI: 10.2519/jospt.2006.36.4.242

Keywords

measurement; neck; spine

Ask authors/readers for more resources

Study Design: This study compared range of motion (ROM) measurements using a cervical range of motion device (CROM) and an optoelectronic system (OPTOTRAK). Objectives: To examine the criterion validity of the CROM for the measurement of cervical ROM on healthy adults. Background: Whereas measurements of cervical ROM are recognized as part of the assessment of patients with neck pain, few devices are available in clinical settings. Two papers published previously showed excellent criterion validity for measurements of cervical flexion/extension and lateral flexion using the CROM. Methods and Measures: Subjects performed neck rotation, flexion/extension, and lateral flexion while sitting on a wooden chair. The ROM values were measured by the CROM as well as the OPTOTRAK. Results: The cervical rotational ROM values using the CROM demonstrated a good to excellent linear relationship with those using the OPTOTRAK: right rotation, r = 0.89 (95% confidence interval, 0.81-0.94), and left rotation, r = 0.94 (95% confidence interval, 0.90-0.97). Similar results were also obtained for flexion/extension and lateral flexion ROM values. Conclusion: The CROM showed excellent criterion validity for measurements of cervical rotation. We propose using ROM values measured by the CROM as outcome measures for patients with neck pain.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available