4.8 Article

Crosslinked chitosan implants as potential degradable devices for brachytherapy: In vitro and in vivo analysis

Journal

JOURNAL OF CONTROLLED RELEASE
Volume 111, Issue 3, Pages 281-289

Publisher

ELSEVIER
DOI: 10.1016/j.jconrel.2005.12.014

Keywords

degradable implant; brachytherapy; chitosan; in vivo degradation; I-131-norcholesterol

Ask authors/readers for more resources

Compared with conventional external beam radiation, brachytherapy offers a superior therapeutic regimen. However, some major constraints are associated with its implementation, including the need of complicated procedures for device placement and removal. The purpose of this study was to examine whether crosslinked chitosan (Ct) implants could serve as potential biodegradable devices for brachytherapy. Ct was reacted with increasing amounts of glutaraldehyde to obtain hydrogels with different crosslinking densities, which were characterized chemically, thermally and mechanically. The effect of the dialysis medium conditions (ionic strength, osmolarity and pH) on the gel hydration and in vivo degradation was assessed. Two types of implants, slow and fast degrading gel (SDG and FDG, respectively), were prepared and implanted with or without Sudan Black (SB) in the rat. While SDG withstood for over a month, the FDG degraded within two weeks after implantation. The release kinetics of SB from the hydrogels verified their in vivo degradation properties. The incorporation of the radioactive compound I-131-norcholesterol (I-131-NC) into the SDG altered the degradation kinetics of the gel as reflected by the release kinetics of the radioactive marker. Eighty percent of I-131-NC was released within a month after implantation, after which time, radioactivity was detected in the regional lymph nodes. Histological examination of the tissues surrounding the implants demonstrated negligible tissue response to the implants, when compared to biodegradable surgical sutures. It is concluded that hydrogels made of crosslinked Ct are potential novel, safe, degradable devices for brachytherapy. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available