4.7 Article

A miniaturized electrochemical system with a novel polyelectrolyte reference electrode and its application to thin layer electroanalysis

Journal

SENSORS AND ACTUATORS B-CHEMICAL
Volume 115, Issue 1, Pages 212-219

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2005.09.005

Keywords

polyelectrolyte reference electrode; microfluidic chip; thin layer cell; salt bridge; differential pulse voltammetry (DPV)

Ask authors/readers for more resources

Miniaturized liquid junction reference electrodes with short- and long-term stabilities were fabricated by using photopolymerization of a patterned polyelectrolytic hydrogel, poly diallyldimethylammonium chloride (pDADMAC). Internal and external solutions were separated by the polyelectrolytic hydrogel barrier, through which Cl- ions carried the ionic current. The impedance spectrum of the developed hydrogel-based salt bridge was flat at 40 k Omega, indicating negligible capacitance. Most of thin film-type reference electrodes patterned on microfluidic chips suffer from a short lifespan and complicated structure, which seriously antagonize the advantages of electrochemical detection. The novel reference electrode system using a polyelectrolyte salt bridge on a microfluidic chip maintained a reproducible potential of 19.3 +/- 6 mV versus a commercial Ag/AgCl reference electrode over 30 h. Anionic interferents such as Br-, I-, and S2- ions had no influence on the performance of the reference electrode. In particular, the dependence on the Cl- ion concentration was substantially suppressed by separating the Ag/AgCl wire from the sample solution with the salt bridge. A 10 mu m-thick cell with a polyelectrolyte salt bridge on a microfluidic chip showed the typical voltammetric behavior of thin layer cells that was confirmed by cyclic voltammetry (CV). This chip-bascd thin layer cell was exploited to determine the concentration of dopamine by differential pulse voltammetry (DPV) as well. A linear current response was observed down to 5 mu M dopamine and the limit of detection was calculated to 2 mu M. (c) 2005 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available