4.2 Article

Development of an immunochromatographic lateral flow test strip for detection of β-adrenergic agonist Clenbuterol residues

Journal

JOURNAL OF IMMUNOLOGICAL METHODS
Volume 312, Issue 1-2, Pages 27-33

Publisher

ELSEVIER
DOI: 10.1016/j.jim.2006.02.017

Keywords

Clenbuterol; test strip; colloidal gold; immunochromatography; rapid test

Ask authors/readers for more resources

A rapid immunochromatographic lateral flow test strip was developed in a competitive format with the gold-conjugated monoclonal antibody to specifically determine the residues of Clenbuterol (CL), a beta-adrenergic agonist. The test strip is made up of a sample pad, a conjugate reagent pad, a test membrane containing a control line and a test line, and an absorbent pad. CL standard samples of 0, 0.1, 0.3, 0.9, 2.7, 8.1 ng/ml in swine urine were determined by the test strip. It was shown that detection limit of the test strip was as low as 0.1 ng/ml of CL and that the half of maximal inhibition concentration (IC50) in relative optical density was calculated to be 1.78 +/- 0.17 ng/ml under an optical density scanner. The sensitivity by eye measurement was 1.0 ng/ml. It takes 10 min to accomplish a test. Parallel analysis of urine samples from pigs fed with CL showed comparable results obtained from the test strip and GC-MS. Therefore, the test strip is very useful as a screening method for quantitative, semi-quantitative or qualitative detection of CL residues in swine urine. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available