Amperometric sensor for heparin: Sensing mechanism and application in human blood plasma analysis

Journal

ELECTROANALYSIS
Volume 18, Issue 13-14, Pages 1329-1338

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200603533

Keywords

heparin; amperometry; PVC membrane electrode; sensing mechanism; human blood plasma

Ask authors/readers for more resources

Voltammetric measurements of heparin at a rotating glassy carbon (GQ electrode coated with a polyvinylchloride membrane are reported. A spin-coating technique is used to prepare thin membranes (20 - 40 mu m) with a composition of 25% (w/w) PVC, 1,1'-dimethylferrocene as a reference electron donor for the GC I membrane interface, nitrophenyl octyl ether (o-NPOE) or bis(2-ethylhexyl) sebacate (DOS) as a plasticizer, and hexadecyltrimethylammonium tetrakis(4-chlorophenyl) borate (HTMATPBCl) or tridodecy1methylammonium tetrakis(4-chlorophenyl) borate (TDMATPBCI) as a background electrolyte. It is shown that the electrodes coated with either the HTMA'/oNPOE (DOS) or TDMA(+)/(O)-NPOE (DOS) membrane provide a comparable amperometric response towards heparin (1 - 10 U mL(-1)) in the aqueous solution of 0.1 M LiCl. However, only the membranes formulated with TDMATPBCl can be used for an amperometric assay of heparin in human blood plasma with a detection limit of 0.2 U mL(-1). Effects of membrane composition, heparin concentration, rotation speed and sweep rate on the voltammetric behavior of heparin provide some insight into the sensing mechanism. Theoretical analysis of the amperometric response is outlined, and the numeric simulation of the voltammetric behavior is presented.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available