4.5 Article

Cardiac fat navigator-gated steady-state free precession 3D magnetic resonance angiography of coronary arteries

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 56, Issue 1, Pages 210-215

Publisher

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.20938

Keywords

coronary magnetic resonance angiography; cardiac fat navigator echo; steady-state free precession; magnetization preparation; cardiac motion

Ask authors/readers for more resources

Motion artifacts and the lack of accurate detection of cardiac motion present a major challenge for high-resolution cardiac MRI. Recently a multidimensional cardiac fat navigator was proposed to provide a fast and direct measurement of bulk cardiac motion. The objective of this study was to demonstrate the feasibility of employing the cardiac fat navigator in balanced steady-state free precession (SSFP) free-breathing 3D coronary MRA (CMRA). The cardiac fat navigator echo is optimized to provide both motion monitoring and epicardial fat suppression. Steady-state magnetization preparation, which is needed for SSFP CMRA, is optimized by comparing three preparation schemes: alpha/2, linear ramp with 20 RF pulses (20LR), and Kaiser ramp with six RF pulses (6KR). The present preliminary human study shows that the 6KR preparation provides better image quality than both the alpha/2 (P < 0.0025) and the 20LR preparations (P < 0.025) for free-breathing SSFP 3D CMRA (N = 11).

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available