4.3 Article

Transverse tripolar spinal cord stimulation: Results of an international multicenter study

Journal

NEUROMODULATION
Volume 9, Issue 3, Pages 192-203

Publisher

WILEY
DOI: 10.1111/j.1525-1403.2006.00060.x

Keywords

spinal cord stimulation; SCS transverse tripole; neuromodulation; FBS; neuropathic pain; failed back syndrome

Ask authors/readers for more resources

Experienced neurosurgeons at eight spinal cord stimulation centers in the United States, Canada, and Europe participated in a study from 1997 to 2000 investigating the safety, performance, and efficacy of a Transverse Tripolar Stimulation (TTS) system invented at the University of Twente, the Netherlands. This device was proposed to improve the ability of spinal cord stimulation to adequately overlap paresthesia to perceived areas of pain. Fifty-six patients with chronic, intractable neuropathic pain of the trunk and/or limbs more than three months' duration (average 105 months) were enrolled with follow-up periods at 4, 12, 26, and 52 weeks. All patients had a new paddle-type lead implanted with four electrodes, three of them aligned in a row perpendicular to the cord. Fifteen of these patients did not undergo permanent implantation. Of the 41 patients internalized, 20 patients chose conventional programming using an implanted pulse generator to drive four electrodes, while 21 patients chose a tripole stimulation system, which used radiofrequency power and signal transmission and an implanted dual-channel receiver to drive three electrodes using simultaneous pulses of independently variable amplitude. On average, the visual analog scale scores dropped more for patients with TTS systems (32%) than for conventional polarity systems (16%). Conventional polarity systems were using higher frequencies on average, while usage range was similar. Most impressive was the well-controlled steering of the paresthesias according to the dermatomal topography of the dorsal columns when using the TTS-balanced pulse driver. The most common complication was lead migration. While the transverse stimulation system produced acceptable outcomes for overall pain relief, an analysis of individual pain patterns suggests that it behaves like spinal cord stimulation in general with the best control of extremity neuropathic pain. This transverse tripole lead and driving system introduced the concept of electrical field steering by selective recruitment of axonal nerve fiber tracts in the dorsal columns.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available