Journal
MEDICAL PHYSICS
Volume 33, Issue 7, Pages 2632-2641Publisher
AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2207367
Keywords
motion artifacts; 4D PET; respiratory motion; deformable registration; b-spline
Ask authors/readers for more resources
Respiratory motion artifacts in positron emission tomography (PET) imaging can alter lesion intensity profiles, and result in substantially reduced activity and contrast-to-noise ratios (CNRs). We propose a corrective algorithm, coined retrospective stacking (RS), to restore image quality without requiring additional scan time. Retrospective stacking uses b-spline deformable image registration to combine amplitude-binned PET data along the entire respiratory cycle into a single respiratory end point. We applied the method to a phantom model consisting of a small, hot vial oscillating within a warm background, as well as to (18)FDG-PET images of a pancreatic and a liver patient. Comparisons were made using cross-section visualizations, activity profiles, and CNRs within the region of interest. Retrospective stacking was found to properly restore the lesion location and intensity profile in all cases. In addition, RS provided CNR improvements up to three-fold over gated images, and up to five-fold over ungated data. These phantom and patient studies demonstrate that RS can correct for lesion motion and deformation, while substantially improving tumor visibility and background noise. (C) 2006 American Association of Physicists in Medicine.