4.6 Article

Trace analysis of hydrogen sulfide by monitoring As(III) at a poly(L-lactide) stabilized gold nanoparticles modified electrode

Journal

ELECTROCHEMISTRY COMMUNICATIONS
Volume 8, Issue 8, Pages 1369-1374

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.elecom.2006.06.014

Keywords

sulfide; arsenite; differential pulse voltammetry; screen-printed electrode

Ask authors/readers for more resources

An indirect electrochemical approach for the determination of sulfide by measuring the inhibited oxidation current of As(III) in HCl medium using a poly(L-lactide) stabilized gold nanoparticles modified screen-printed electrode (designated as PLA-Au-NP/SPE) is described in this work. The PLA-AU(NP)/SPE is applied effectively to detect As(111) in HCl medium and can be tolerable from the interference of Cu, Cd, Fe, Zn, Mn and Ni and hence provides a direct and selective detection method for As(111) in natural waters. An substantial decrease in the oxidation peak current of As(III) is observed in the presence of sulfide due to the formation of As2S3 complex. Surface analysis by XPS together with electrochemical and Raman spectroscopic characterization were used to elucidate the detection mechanism. Under the optimal experimental conditions, a calibration plot with a wide linear range up to 700 mu M was achieved with a detection limit (S/N = 3) of 0.04 mu M. The sensitivity observed in present study is good enough to detect sulfide at levels lower than the current EPA standard. The proposed electrochemical approach was successfully used for the determination of sulfide in acid rain and hot spring water. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available