Journal
ELECTROANALYSIS
Volume 18, Issue 15, Pages 1493-1498Publisher
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200603547
Keywords
electrochemical impedance spectroscopy (EIS); solid-state pH electrode; titanium nitride (TiN); pH-sensitive; nenstian response
Categories
Ask authors/readers for more resources
Titanium nitride was used as pH-sensitive material to fabricate all solid-state pH electrode. The fabrication and the response performance of the pH electrode were described in the paper. The TiN film electrode showed a linear response in the pH range of 2-12 with a near-Nenstian response (-55 mV/pH). The response time was within 1 min, and the electrode had good reproducibility, stability and low sensitivities for different species. Compared with the glass pH electrode, the electrode exhibited some advantages, for example, without activation, rapid response and high mechanical strength. In addition, the electrode performed excellently in a corrosion medium containing F-(1 M). Electrochemical behaviors of TiN electrode in Britton-Robinson buffers were studied with Electrochemical Impedance Spectroscopy (EIS).
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available