4.6 Article

Two cases of ischemia associated with subthalamic nucleus stimulator implantation for advanced Parkinson's disease

Journal

MOVEMENT DISORDERS
Volume 21, Issue 9, Pages 1477-1483

Publisher

WILEY-LISS
DOI: 10.1002/mds.20947

Keywords

Parkinson's disease; deep brain stimulation; subthalamic nucleus; complications; ischemia; microelectrode recording

Ask authors/readers for more resources

Deep brain stimulation is generally a safe and effective method of alleviating motor impairment in advanced-stage Parkinson's disease patients. However, adverse events of surgery have been noted, such as hemorrhage, infection, seizures, and device failure. In this report, we describe 2 cases of the unusual adverse event of ischemia associated with subthalamic nucleus stimulator implantation. We present the intraoperative neurological symptoms, microelectrode recording data, imaging findings, and other correlated events. In the first case, the clinical effects of ischemia were evident intraoperatively and coincided with silence during microelectrode recording from the ischemic region. In the second case, the timing of the ischemic event could not be determined precisely but also was associated with a difficult mapping. Subcortical ischemia may be an underrecognized event that confounds neurophysiological mapping of deep brain structures and affects clinical outcomes. (c) 2006 Movement Disorder Society.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available