4.2 Article

Mean platelet volume measurement, EDTA or citrate?

Journal

HEMATOLOGY
Volume 11, Issue 5-6, Pages 317-319

Publisher

TAYLOR & FRANCIS LTD
DOI: 10.1080/10245330600954163

Keywords

MPV; anticoagulant; EDTA; citrate; automated counters

Categories

Ask authors/readers for more resources

Most laboratories use EDTA for anticoagulation of whole blood prior to automated cell counting but due to platelet swelling, mean platelet volume (MPV) values may increase with its use. MPV changes may be less with acid citrate based anticoagulation. As MPV is a marker of platelet function and its precise measurement is important in a number of clinical situations, this study was performed to assess if EDTA and citrate based anticoagulated blood samples can be used interchangeably for MPV measurement. In this cross sectional descriptive study, EDTA and citrate based anticoagulated blood samples of the same patients were assessed by auto-analyzer within 1 h of sampling. In the 61 evaluated patients, there was a close correlation between MPV as measured by EDTA and citrate, but mean MPV measured from EDTA samples was 0.66 fL (9%) more than citrate. There was also a significant negative correlation between platelets count and MPV by both methods. The results of our study reveal that MPV can be measured accurately by both methods of anticoagulation; EDTA and citrate if analysis be performed within 1 h of sampling.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available