4.2 Article

Method development and validation of a rapid determination of ropinirole in tablets by LC-UV

Journal

CHROMATOGRAPHIA
Volume 64, Issue 7-8, Pages 459-461

Publisher

SPRINGER HEIDELBERG
DOI: 10.1365/s10337-006-0045-1

Keywords

column liquid chromatography; tablet analysis; quality control; ropinirole

Ask authors/readers for more resources

A reversed-phase high-performance liquid chromatographic (HPLC) method with UV detection was developed and validated for the determination of ropinirole (ROP) in tablets. The assay utilized UV detection at 250 nm and a Luna CN column (250 x 4.6 mm I.D, 5 mu m). The mobile phases were comprised of acetonitrile: 10 mM nitric acid (pH 3.0) (75:25, v/v). Validation experiments were performed to demonstrate linearity, accuracy, precision, limit of quantitation (LOQ), limit of detection (LOD), and robustness. The method was linear over the concentration range of 0.5-10.0 mu g mL(-1). The method showed good recoveries (99.75-100.20%) and the relative standard deviations of intra and inter-day assays were 0.38-1.69 and 0.45-1.95%, respectively. The method can be used for quality control assay of ropinirole.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available