4.7 Article

Construction and calibration of an instrument for three-dimensional ion imaging

Journal

JOURNAL OF CHEMICAL PHYSICS
Volume 125, Issue 13, Pages -

Publisher

AIP Publishing
DOI: 10.1063/1.2217742

Keywords

-

Ask authors/readers for more resources

We describe a new instrument based on a delay-line detector for imaging the complete three-dimensional velocity distribution of photoionized products from photoinitiated reactions. Doppler-free [2+1] resonantly enhanced multiphoton ionization (REMPI) of H and D atoms formed upon photolysis of HBr and DBr in the range 203 nm <=lambda(photolysis)<= 243 nm yields radial speeds measured to be accurate within 1% of those calculated. The relative speed resolution is about 5% and limited by photoionization recoil broadening. A relative speed resolution of 3.4% is obtained for [3+1] REMPI, which minimizes the ionization recoil. We also determine the branching ratio between ground-state and spin-orbit-excited product channels and their associated anisotropies. We find that DBr photolysis dynamics differs slightly from its HBr counterpart. (c) 2006 American Institute of Physics.