4.5 Article

Spinal cord monitoring in scoliosis surgery using an epidural electrode. Results of a prospective, consecutive series of 191 cases

Journal

SPINE
Volume 31, Issue 22, Pages 2614-2623

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.brs.0000240642.28495.99

Keywords

somatosensory evoked potential; neurogenic mixed evoked potential; scoliosis surgery; functional neurologic integrity; epidural electrode

Ask authors/readers for more resources

Study Design. Retrospective analysis of a prospectively accrued series of 191 consecutive patients who underwent intraoperative neurophysiologic monitoring during scoliosis corrective surgery. Objectives. To compare the monitoring outcome of idiopathic and neuromuscular scoliosis. To demonstrate the usefulness of the epidural electrode. To report sensitivity and specificity of the monitoring method employed at a single institution. Summary of Background Data. Reports in the literature emphasized the difficulty to obtain data in neuromuscular patients. Multimodality spinal cord monitoring has been recommended. Despite their still debated composition, neurogenic motor-evoked potentials have proven their validity in clinical practice. Methods. Somatosensory and neurogenic evoked potentials were attempted in all patients presenting for scoliosis correction between 1999 and 2005. Study patients were divided into 3 groups: group 1, idiopathic; group 2, neuromuscular; and group 3, miscellaneous origins. Results. The use of the epidural electrode demonstrated significant usefulness in the ability of monitoring otherwise nonmonitored patients, especially in group 2. Inability to obtain any evoked potentials occurred in 4 cases (2.1%). Five cases were found to be true positives. An adapted and rapid intervention permitted to avoid new postoperative deficit in all cases. There was no instance of false-negative data. The overall method sensitivity was 100%, and specificity was 52.69%. Conclusions. The use of a single epidural electrode allowing somatosensory evoked potentials recording and spinal cord stimulation alternately is a safe and valid method of intraoperative monitoring.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available