4.4 Article

Quantum cascade laser-based integrated cavity output spectroscopy of exhaled nitric oxide

Journal

APPLIED PHYSICS B-LASERS AND OPTICS
Volume 85, Issue 2-3, Pages 445-452

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00340-006-2365-0

Keywords

-

Ask authors/readers for more resources

A nitric oxide (NO) sensor employing a thermoelectrically cooled, continuous-wave, distributed feedback quantum cascade laser operating at 5.47 mu m (1828 cm(-1)) and off-axis integrated cavity output spectroscopy was used to measure NO concentrations in exhaled breath. A minimum measurable concentration (3 sigma) of 3.6 parts-per-billion by volume (ppbv) of NO with a data-acquisition time of 4 s was demonstrated. Five prepared gas mixtures and 15 exhaled breath samples were measured with both the NO sensor and for intercomparison with a chemiluminescence-based NO analyzer and were found to be in agreement within 0.6 ppbv. Exhaled NO flow-independent parameters, which may provide diagnostic and therapeutic information in respiratory diseases where single-breath measurements are equivocal, were estimated from end-tidal NO concentration measurements collected at various flow rates. The results of this work indicate that a laser-based exhaled NO sensor can be used to measure exhaled nitric oxide at a range of exhalation flow rates to determine flow-independent parameters in human clinical trials.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available