4.8 Article

Simultaneous mass analysis of positive and negative ions using a dual-polarity time-of-flight mass spectrometer

Journal

ANALYTICAL CHEMISTRY
Volume 78, Issue 22, Pages 7729-7734

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/ac061213v

Keywords

-

Ask authors/readers for more resources

Positive and negative ions produced from matrix-assisted laser desorption/ionization (MALDI) were simultaneously measured using a newly developed dual-polarity time-of-flight mass spectrometer. This instrument is effective not only for express and comprehensive mass analysis but also for studying the ionization mechanisms of biomolecules. It comprises two identical time-of-flight mass analyzers located symmetrically about a MALDI ion source. The ion optics are arranged to be able to extract positive and negative ions synchronously with equal efficiency to each corresponding mass analyzer. Mass spectra of various proteins with molecular weights as large as that of myoglobin monomer and dimer were obtained. The spectral patterns obtained in this work are approximately mirror images with opposite polarities.