4.2 Article

Validation study of fractional exhaled nitric oxide measurements using a handheld monitoring device

Journal

JOURNAL OF ASTHMA
Volume 43, Issue 10, Pages 731-734

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/02770900601031045

Keywords

asthma; exhaled nitric oxide; airway inflammation; point of care

Ask authors/readers for more resources

We tested reproducibility of exhaled nitric oxide (FENO) and inter-operator handling when measured with a handheld device, NIOX MINO(R). We enrolled 20 volunteers using a priori goals of acceptable reproducibility to be mean within-subject standard deviation less than 3 parts per billion (ppb) for FENO measurements less than 30 ppb, and mean coefficient of variation less than 10% for FENO measurements more than 30 ppb. Seventeen subjects with measurements less than 30 ppb displayed a mean standard deviation of 1.15, and 3 subjects with FENO more than 30 ppb had a mean coefficient of variation of 2.4%. We conclude that NIOX MINO demonstrates excellent reproducibility for all ranges of FENO.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available