4.7 Article

A PVC plasticized membrane sensor for nickel ions

Journal

MICROCHIMICA ACTA
Volume 156, Issue 3-4, Pages 283-287

Publisher

SPRINGER WIEN
DOI: 10.1007/s00604-006-0572-3

Keywords

nickel-selective sensor; PVC matrix; thiophene-derivative Schiff base; potentiometry

Ask authors/readers for more resources

A new PVC membrane sensor, which is highly selective towards Ni (II) ions, has been developed using a thiophene-derivative Schiff base as the ionophore. The best performance was exhibited by the membrane having the composition percentage ratio of 5:3:61:31 (ionophore:NaTPB:DBP:PVC) (w/w), where NaTPB is the anion excluder, sodium tetraphenylborate and DBP is the plasticizing agent (dibutyl phthalate). The membrane exhibited a good Nernstian response for nickel ions over the concentration range of 1.0 x 10(-1)-5.0 x 10(-6) M (limit of detection is 1.8 x 10(-6) M) with a slope of 29.5 +/- 1.0 mV per decade of activity. It has a fast response time of < 20 s and can be used for a period of 4 months with good reproducibility. The sensor is suitable for use in aqueous solutions of a wide pH range of 3.2-7.9. The sensor shows high selectivity to nickel ions over a large number of mono-, bi- and trivalent cations. It has been successfully used as an indicator electrode in the potentiometric titration of nickel ions against EDTA and also for direct determination of nickel content in real samples - wastewater samples from electroplating industries and Indian chocolates.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available