4.5 Article

Laboratory calibration of the Extreme-Ultraviolet Imaging Spectrometer for the Solar-B satellite

Journal

APPLIED OPTICS
Volume 45, Issue 34, Pages 8689-8705

Publisher

OPTICAL SOC AMER
DOI: 10.1364/AO.45.008689

Keywords

-

Categories

Ask authors/readers for more resources

The laboratory end-to-end testing of the Extreme-Ultraviolet Imaging Spectrometer (EIS) for the Solar-B satellite is reported. A short overview of the EIS, which observes in two bands in the extreme-ultraviolet wavelength range, is given. The calibration apparatus is described, including details of the light sources used. The data reduction and analysis procedure are outlined. The wavelength calibration using a Penning source to illuminate the aperture fully is presented. We discuss the aperture determination using a radiometrically calibrated hollow-cathode-based source. We then give an account of the predicted and measured efficiencies from consideration of the efficiencies of individual optical elements in first order, an account of efficiencies out of band when radiation incident in one band is detected in the other, and efficiencies in multiple orders. The efficiencies measured in first order for in band and out of band are compared with the predictions and the sensitivity, and its uncertainties are derived. Application of the radiometric calibration is discussed. (c) 2006 Optical Society of America.