Fast gas chromatographic/mass spectrometric determination of diuretics and masking agents in human urine -: Development and validation of a productive screening protocol for antidoping analysis

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1135, Issue 2, Pages 219-229

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2006.09.034

Keywords

fast GC/MS; diuretics; masking agents; doping control; urine screening; ECNI

Ask authors/readers for more resources

An analytical procedure was developed for the fast screening of 16 diuretics (acetazolamide, althiazide, amiloride, bendroflumethiazide, bumetanide, canrenoic acid, chlorthalidone, chlorthiazide, clopamide, ethacrynic acid, furosemide, hydrochlorthiazide, hydroflumethiazide, indapamide, triamterene, trichlormethiazide) and a masking agent (probenecid) in human urine. The whole method involves three analytical steps, including (1) liquid/liquid extraction of the analytes from the matrix, (2) their reaction with methyl iodide at 70 degrees C for 2 h to form methyl derivatives, (3) analysis of the resulting mixture by fast gas chromatography/electron impact mass spectrometry (fast GC/EI-MS). The analytical method was validated by determining selectivity, linearity, accuracy, intra and inter assay precision, extraction efficiencies and signal to noise ratio (S/N) at the lowest calibration level (LCL) for all candidate analytes. The analytical performances of three extraction procedures and five combination of derivatization parameters were compared in order to probe the conditions for speeding up the sample preparation step. Limits of detection (LOD) were evaluated in both EI-MS and ECNI-MS (electron capture negative ionization mass spectrometry) modes, indicating better sensitivity for most of the analytes using the latter ionization technique. The use of short columns and high carrier gas velocity in fast GC/MS produced efficient separation of the analytes in less than 4 min, resulting in a drastic reduction of the analysis time, while a resolution comparable to that obtained from classic GC conditions is maintained. Fast quadrupole MS electronics allows high scan rates and effective data acquisition both in scan and selected ion monitoring modes. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available