4.6 Article

Development and validation of a simple HPLC method for simultaneous in vitro determination of amoxicillin and metronidazole at single wavelength

Journal

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
Volume 43, Issue 1, Pages 325-329

Publisher

ELSEVIER
DOI: 10.1016/j.jpba.2006.06.002

Keywords

HPLC; amoxicillin; metronidazole; simultaneous determination; sustained release; method validation

Ask authors/readers for more resources

A simple, rapid, sensitive and robust reversed phase-HPLC method was developed and validated to measure simultaneously the amount of amoxicillin and metronidazole at single wavelength (254 nm) in order to assess drug release profiles and drug-excipients compatibility studies for a new floating-sustained release tablet formulation and its subsequent stability studies. An isocratic elution of filtered sample was performed on C I g column with buffered mobile phase (pH 4.0) and UV detection at 254 nm. Quantification was achieved with reference to the external standards. The linearity for concentrations between 0.15 and 600 mu g/ml for amoxicillin and 0.13 and 300 mu g/ml for metronidazole were established. Intra and inter-day precision were less than 2.5%. The limits of detection (LOD) and quantification were 0.05 and 0.15 mu g/ml for amoxicillin and 0.10 and 0.13 mu g/ml for metronidazole. The determination of the two active ingredients was not interfered by the excipients of the products. Samples were stable in the release media (37 degrees C) and the HPLC injector at least for 12 h. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available