4.5 Article

Stabilization of an optical microscope to 0.1 nm in three dimensions

Journal

APPLIED OPTICS
Volume 46, Issue 3, Pages 421-427

Publisher

OPTICAL SOC AMER
DOI: 10.1364/AO.46.000421

Keywords

-

Categories

Ask authors/readers for more resources

Mechanical drift is a long-standing problem in optical microscopy that occurs in all three dimensions. This drift increasingly limits the resolution of advanced surface-coupled, single-molecule experiments. We overcame this drift and achieved atomic-scale stabilization (0.1 nm) of an optical microscope in 3D. This was accomplished by measuring the position of a fiducial mark coupled to the microscope cover slip using back-focal-plane (BFP) detection and correcting for the drift using a piezoelectric stage. Several significant factors contributed to this experimental realization, including (i) dramatically reducing the low frequency noise in BFP detection, (ii) increasing the sensitivity of BFP detection to vertical motion, and (iii) fabricating a regular array of nanometer-sized fiducial marks that were firmly coupled to the cover slip. With these improvements, we achieved short-term (1 s) stabilities of 0.11, 0.10, and 0.09 nm (rms) and long-term (100 s) stabilities of 0.17, 0.12, and 0.35 nm (rms) in x, y, and z, respectively, as measured by an independent detection laser. (c) 2007 Optical Society of America

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available