Information loss in visual assessments of medical images

Journal

EUROPEAN JOURNAL OF RADIOLOGY
Volume 61, Issue 2, Pages 362-366

Publisher

ELSEVIER IRELAND LTD
DOI: 10.1016/j.ejrad.2006.09.009

Keywords

information theory; information loss; contrast-detail (C-D) analysis; visual assessment

Ask authors/readers for more resources

We applied information theory to quantify information losses in assessing contrast-detail (C-D) analysis. Images of a C-D phantom were acquired with a flat panel detector (FPD) and a computed radiography (CR) by changing surface entrance doses. Six phantom radiographs (FPD: five images; CR: one image) were prepared for visual evaluations. Thirteen radiographers and two radiologists participated in the observation test. Delectability was defined as the shortest length of the cylinders of which border the observers could recognize from the background, and was recorded using row number. Information content was defined as the entropy Sigma p(i)log(l/p(i)) with detection probabilities pi, which were calculated from distribution of detection rate of the ith column. Information loss, in unit of bits, was calculated as the difference between information obtained and information content when all the columns were detected. The information losses decreased with the increase in cylinder diameters and with the increase in surface entrance dose. Because the information loss varies depending on distribution of detection rate, this method of using the information theory was expected to be more sensitive in evaluating the C-D image quality than using the averaged values of delectability. (c) 2006 Elsevier Ireland Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available