Journal
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume 571, Issue 1-2, Pages 26-29Publisher
ELSEVIER SCIENCE BV
DOI: 10.1016/j.nima.2006.10.020
Keywords
PET; continuous scintillator; PSPMT
Ask authors/readers for more resources
We have constructed a small animal PET with four identical detector modules, each consisting of a continuous LYSO crystal attached to a Position Sensitive Photomultiplier Tube (PSPMT). The dimensions of the continuous crystal are 50 x 50 mm 2 and 10 mm thickness. The modules are separated 11 cm between each other in the scanner. In this paper we discuss the method used for the calibration of the camera for this special system with continuous detectors. We also present the preliminary values for the main performance parameters such as spatial and energy resolution, and sensitivity of the system. (c) 2006 Elsevier B.V. All rights reserved.