Organ and effective doses in infants undergoing upper gastrointestinal (UGI) fluoroscopic examination

Journal

MEDICAL PHYSICS
Volume 34, Issue 2, Pages 703-710

Publisher

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2426405

Keywords

pediatric radiology; fluoroscopy; upper gastrointestinal; organ dose; effective dose; tomographic phantom

Funding

  1. NIBIB NIH HHS [R01 EB00267] Funding Source: Medline
  2. NICHD NIH HHS [R01 HD38932] Funding Source: Medline

Ask authors/readers for more resources

To provide more detailed data on organ and effective doses in digital upper gastrointestinal (UGI) fluoroscopy studies of newborns and infants, the present study was conducted employing the time-sequence videotape-analysis technique used in a companion study of newborn and infant voiding cystourethrograms (VCUG). This technique was originally pioneered [O. H. Suleiman, J. Anderson, B. Jones, G. U. Rao, and M. Rosenstein, Radiology 178, 653-658 (1991)] for adult UGI examinations. Individual video frames were analyzed to include combinations of field size, field center, x-ray projection, image intensifier, and magnification mode. Additionally, the peak tube potential and the mA or mAs values for each segment/subsegment or digital photospot were recorded for both the fluoroscopic and radiographic modes of operation. The data from videotape analysis were then used in conjunction with a patient-scalable newborn tomographic computational phantom to report both organ and effective dose values via Monte Carlo radiation transport. The study includes dose estimates for five simulated UGI examinations representative of patients ranging from three to six months of age. Effective dose values for UGI examinations ranged from 1.17 to 6.47 mSv, with a mean of 3.14 mSv and a large standard deviation of 2.15 mSv. The colon; lungs, stomach, liver, and esophagus absorbed doses in sum were found to constitute between 63 and 75% of the effective dose in these UGI studies. Representing 23-30% of the effective dose, the lungs were found to be the most significant organ in the effective dose calculation. Approximately 80-95% of the effective dose is contributed by the dynamic fluoroscopy segments with larger percentages found in longer studies. The mean effective dose for newborn UGI examinations was not found to be statistically different from that seen in newborn VCUG examinations. (c) 2007 American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available