4.5 Article

Comparison of the performance of different reversed-phase columns for liquid chromatography separation of 11 pollutant phenols

Journal

JOURNAL OF SEPARATION SCIENCE
Volume 30, Issue 5, Pages 699-707

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.200600301

Keywords

pollutant phenols; reversed-phase columns

Ask authors/readers for more resources

A systematic optimization of the HPLC separation of a mixture containing 11 pollutant phenols (PPs) using a Hypersil ODS (250 mm x 4.6 mm, 5 mu m) column and UV-DAD detection has been carried out. The binary mobile phases used were obtained by mixing 50 mM phosphate (pH = 3.0) and methanol, ACN, or THF as organic modifiers. After selecting ACN as an organic modifier, the effects of pH and temperature on PPs separation were studied. A mobile phase of 50 mM acetate (pH = 5.0) - ACN (60:40 v/v) at 50 degrees C allowed the separation of 11 phenols but not to baseline in 17 min. To improve the performance of this separation, the following RP columns were tested: Luna C18 (2), Purospher C18, Synergi C12, Synergi Fusion C18, Gemini C18, Luna Cyano, Lichrospher C8, and Envirosep-PP (polymeric). In all the cases, the performance (analysis time, retention, selectivity, resolution, asymmetry factors, and efficiency) was evaluated. A further reoptimization of the mobile phase was carried out for all the columns by studying the ACN content and pH, with the aim of improving the above-mentioned separations and selecting the most suitable one for PPs analysis.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available